Manufacturer: Sun International Co., Ltd., Japan   |   Authorized Distributor in India: Akira Analytical Solutions Pvt. Ltd.
HiQ Sil chromatography columns

HiQ Sil Columns

Japanese silica technology · Indian technical support

HiQ Sil Technical Resources

Use the jump links to reach each topic. These recommendations are general; always follow the current manufacturer instructions for the exact column and method.

  1. How to Select a HiQ Sil Column
  2. Reversed-Phase Column Selection
  3. Normal-Phase and Polar Phase Selection
  4. Particle-Size Selection
  5. Pore-Size Selection
  6. Column Installation
  7. Column Conditioning
  8. Mobile-Phase Preparation
  9. Column Flushing
  10. Column Storage
  11. Guard Columns
  12. High Backpressure Troubleshooting
  13. Peak Tailing Troubleshooting
  14. Peak Fronting
  15. Split Peaks
  16. Retention-Time Changes
  17. Loss of Efficiency
  18. Ghost Peaks
  19. Column Lifetime
  20. Column Regeneration
  21. Method Transfer
  22. Safety and Good Laboratory Practice

How to Select a HiQ Sil Column

Start with analyte polarity and ionization, then consider the required retention mechanism, selectivity, particle size, pore size, internal diameter, length and instrument pressure capability. Compare alternative phases when critical compounds co-elute.

Reversed-Phase Column Selection

C18 generally gives the strongest hydrophobic retention. C8 and C4 reduce retention and can shorten run time for highly hydrophobic compounds. Phenyl adds π–π interactions and can change aromatic selectivity.

Normal-Phase and Polar Phase Selection

NH2, CN and bare silica provide polar surface chemistry. Normal-phase methods require careful control of solvent water content because small moisture changes can alter retention.

Particle-Size Selection

Three-micrometre particles can provide higher efficiency but typically create more backpressure. Five-micrometre particles are widely used for routine analytical work. Ten-micrometre particles provide lower pressure and are suited to conventional methods where maximum efficiency is not required.

Pore-Size Selection

The catalogue lists 100 Å and 120 Å materials. Pore size influences accessible surface area and analyte diffusion. Choose according to analyte size and the validated application.

Column Installation

Confirm flow direction, use compatible low-dead-volume fittings, avoid overtightening and increase flow gradually. A poor connection can cause leaks, peak splitting or excess dead volume.

Column Conditioning

Equilibrate with a fully compatible mobile phase until pressure, baseline and retention are stable. Gradient methods may require several column volumes for complete re-equilibration.

Mobile-Phase Preparation

Use HPLC-grade solvents, accurate volumetric preparation, filtration where required and adequate degassing. Prevent buffer precipitation when changing to high-organic solvents.

Column Flushing

Flush according to the stationary phase and contaminants present. First remove salts with a compatible aqueous-organic mixture, then use stronger organic solvent if needed. Never switch abruptly between immiscible solvents.

Column Storage

Remove buffers and salts, store in a compatible solvent and cap both ends. Record the storage solvent so the next user can recondition safely.

Guard Columns

Guard cartridges intercept particulates and strongly retained contaminants before they reach the analytical bed. Replace them when pressure rises or chromatographic performance changes. A compatible adaptor is required.

High Backpressure Troubleshooting

Check the system without the column, inspect tubing and filters, reverse-check the inlet frit only when manufacturer guidance permits, and consider particulate blockage, buffer precipitation, viscous mobile phase or a contaminated guard cartridge.

Peak Tailing Troubleshooting

Possible causes include secondary interactions, inappropriate pH, sample overload, contamination, dead volume and poor fittings. Adjust one factor at a time and compare a system-suitability standard.

Peak Fronting

Common causes include sample overload, strong injection solvent, column-bed damage or poor focusing. Reduce load and match the sample solvent more closely to the initial mobile phase.

Split Peaks

Investigate sample-solvent mismatch, poorly seated fittings, partial frit blockage, injection problems and possible void formation.

Retention-Time Changes

Check mobile-phase composition, temperature, flow-rate accuracy, equilibration, buffer pH, solvent evaporation and column history.

Loss of Efficiency

Potential causes include contamination, voids, excessive injection volume, overload, worn system components and insufficient equilibration.

Ghost Peaks

Check carryover, solvent impurities, contaminated glassware, autosampler wash settings, system contamination and sample-preparation reagents.

Column Lifetime

Lifetime depends on sample cleanliness, mobile phases, pH exposure, pressure, temperature, system cleanliness and maintenance. A fixed injection count should not be promised.

Column Regeneration

Recovery depends on the contaminant and whether the bed is physically intact. Progressive flushing may restore performance, but damaged beds or permanently adsorbed material may not be recoverable.

Method Transfer

Account for internal diameter, length, particle size, flow rate, injection volume, gradient volume and instrument dwell volume. Revalidate system suitability after transfer.

Safety and Good Laboratory Practice

Handle solvents in accordance with safety data sheets, use suitable ventilation and PPE, control waste streams and follow manufacturer documentation and laboratory procedures.